ODYS - CTL 2386
| Basic technical parameters of scanner |
| Scanning angle |
+/- 12o |
| Shape of irradiated surface |
circle, ellipse, square or rectangle |
| Scanning frequency |
(0.1÷1)Hz |
| Control system |
from the keyboard, with built-in microcomputers |
| Optical system |
two moving mirrors and set of mirrors directing laser beams to overlap |
| Aiming beam |
built-in diode laser lighting the operation field by red beam (5mW/650nm) or additionally working as therapy laser (50mW/660nm) |
| Power supply |
single phase (220-230)V, 50Hz |
| Power consumption |
12 W |
| Size |
220mm x 156mm x 100mm |
| Weight |
2.1 kg |
| Saftey |
scanner made according to the international standards IEC 601 and IEC 825 |
| Basic technical parameters of laser |
| Laser type |
diode (semiconductor) laser |
| Wavelength |
(820+/-10)nm, near IR |
| Operation mode |
contnuous wave (c.w.) |
| Radiation power on tissue |
regulated in the range 0÷200mW or 0÷400mW. |
| Control system |
from the keyboard, with built-in microcomputer |
| Displayed functions |
radiation power [mW], energy dose [J], therapy time duration [min., sec.] |
| Application handpiece |
typically ended with lens applicator and additionally provided with easy-to-change scanner tip |
| Additional accessories |
protective spectacles and goggles, robust carrying case, interchangeable and heat-resistant tips with different shapes and sizes |
| Power supply |
single phase, (220-230)V, 50Hz |
| Power consumption |
<18W |
| Controller size |
170mm x 275mm x 110mm |
| Weight |
2kg |
| Safety |
laser made according to the inernational
standards IEC 601.1 i IEC 825. It is equipped with: built-in power
meter, remote interlock connection and key switch. |
|